R.E.B.E.L. EM – Pacemaker Basics

Originally published at R.E.B.E.L. EM on Aug 24, 2017. Reposted with permission.

Follow Salim R. Rezaie at @srrezaie, and Dr. Anand Swaminathan at @EMSwami.


Have you ever been confused by the alphabet soup of pacemakers? This post will serve as a pacemaker basics reference.

Pacemaker Terminology

  • Five letter code system that describes the different pacemaker settings
  • First three letters are most useful in ED management

  • Biventricular Pacing (Spectrumhealth.com)

    Sensing Response

    • Triggered (T): Sensed intrinsic depolarization will result in the pacemaker discharging (this setting not used in current generation pacemakers)
    • Inhibited (I): Sensed intrinsic depolarization will result in inhibition of the pacemaker
    • Dual (D): Dual inhibition of both atrial and ventricular pacing in response to intrinsic ventricular depolarization
    • None (O): Does not trigger or inhibit regardless of the native activity
  • Example #1: DDD
    • Chamber paced: Atrium and ventricle can be paced
    • Chamber sensed: Intrinsic depolarization of the atrium and ventricle is sensed
    • Sensory response: Can inhibit pacemaker in response to an intrinsic ventricular depolarization or trigger in response to an atrial intrinsic depolarization without a ventricular response
  • Example #2: VVI
    • Device has a single lead in the ventricle that senses ventricular activity and can pace the ventricle
    • A ventricular event outside the refractory period will be inhibited

Common Indications for Permanent Pacemaker Placement

  • 3rd degree or advanced 2nd degree (Mobitz II) AV block with any of the following
    • Symptomatic bradycardia
    • Ventricular dysrhythmia resulting from block
    • Symptomatic bradycardia secondary to necessary drugs for dysrhythmia management
  • Symptomatic bradycardia resulting from 2nd degree AV block
  • Chronic bifascicular or trifascicular block with intermittent 3rd degree or 2nd degree AV block
  • Exertional 2nd or 3rd degree AV block

Pacemaker Components

  • Pulse Generator
    • Houses power source (almost always lithium battery)
    • Battery duration: 4-10 years (depends on features of device)
    • Power decreases gradually with lithium battery; not abrupt cessation of activity
  • Electronic Circuitry
  • Leads
    • Single lead: endocardial lead positioned in contact with right ventricle
    • Dual lead: endocardial lead positioned in contact with right atrium and ventricle

Sample Normal EKGs

  • EKG appearance depends on type of pacemaker in use
  • Single Ventricular Lead
    • Narrow “spike” represents pacemaker generated electrical stimulus
    • Followed by LBBB formation QRS complex
    • Note: RBBB formation signifies lead displacement
  • Dual Chamber Pacing
    • Two narrow “spikes” seen representing an atrial stimulus (1st spike) and ventricular stimulus (2nd spike)
    • First spike results in atrial depolarization. Second spike results in ventricular depolarization
    • Ventricular response variable
      • Narrow QRS
        • Conducted from intrinsic atrial activity
        • Atrial spike conducted through AV node and generates QRS
      • Wide QRS
        • Generated by ventricular lead stimulus
    • Absence of paced complexes in a patient with a pacemaker does not always indicate pacemaker dysfunction. It may reflect adequate native conduction.

 

Complications

  • Pacemaker Pocket Infection (researchgate.com)

    Infectious Complications

    • Rare
      • Pocket infection: 2%
      • Bacteremia: 1%
    • Signs/Symptoms: warmth, swelling, erythema, pain, fever
    • Blood cultures: (+) in 20-25% of patients (commonly S. Aureus, S. epidermidis)
    • Can result in pacemaker pocket erosion
    • Treatment
      • IV antibiotics (i.e.vancomycin)
      • Needle aspiration: should only be performed under fluoroscopy to avoid cutting components of the device
        • Cardiology consultation for removal and replacement
  • Thrombophelbitis
    • Common complication (30-50% with some venous obstruction) (Kucher 2011)
    • Symptoms
      • Occur in only 0.5 -3.5% of patients due to collateralization
      • Pain, swelling, venous engorgement of the ipsilateral arm
    • Diagnostic Testing: Duplex sonography
    • Treatment: Anticoagulation
  • Pacemaker Syndrome
    • Definition
      • Loss of AV synchrony and loss of atrial “kick”
      • Occurs when there is intact sinus node function but atrium contracts with tricuspid/mitral valves closed (i.e. during ventricular systole)
    • Symptoms: Syncope/Near-syncope, orthostatic dizziness, exercise intolerance, generalized weakness, palpitations
    • Treatment: replacement of single chamber pacemaker to dual chamber pacemaker

Pacemaker Manufacturer

  • In order to interrogate a pacemaker, the device manufacturer must be known
  • The majority of the time, this information can be determined from the medical record or, the patient will carry a device manufacturer card with all of the information
  • Algorithm to determine manufacturer using X-ray (Jacob 2011)

CaRDIA-X Algorithm

Effect of Magnet on Pacemaker (Core EM Video Link)

  • Magnet application over a pacemaker will lead the pacemaker to default to the “magnet mode”
  • Typically will lead to asynchronous pacing mode
    • Results in a constant paced rate regardless of the native rate or rhythm
    • Either AOO, VOO or DOO
    • The pacemaker will pace either the atrium, ventricle or both but will not sense or have a response to sensing
  • Risk: Asynchronous ventricular pacing can lead to pacemaker-induced ventricular tachycardia

Pacemaker Malfunction

  • Problems with Sensing
    • Undersensing
      • Definition: Pacemaker fails to sense native cardiac activity resulting in asynchronous pacing
      • Can be complete or intermittent
      • Commonly results from exit block
        • Exit block: Failure of an adequate stimulus to depolarize the chamber. Endocardium in contact with pacing lead requires higher stimulation threshold for conduction
        • Exit block results from changes in cardiac tissue over time (i.e. an RV Infarction, progressive cardiomyopathy)
      • EKG Findings
        • Pacing spikes occurring earlier than the programmed rate
        • Spike may not be followed by QRS complex if within the refractory period
        • Presence of spikes within QRS complexes also suggestive
    • Oversensing
      • Definition: Pacemaker mistakes electrical signals as native cardiac activity and thus, pacemaker function is inhibited
      • Sources: Large P or T wave, pectoralis muscle contraction, cell phone signal (typically when on ipsilateral ear)
  • Problems with Pacing
    • Output Failure
      • Definition: No paced stimulus is generated from the device resulting in decreased or absent pacemaker function
      • Causes: Oversensing, Lead displacement, Fracture of pacing wires
    • Failure to Capture
      • Definition: Electrical stimulus does not result in depolarization of the myocardium (no QRS complex generated)
      • Causes: Exit block, myocardial infarction, lead displacement, wire fracture, electrolyte abnormalities

Failure to Capture (LITFL)

  • Pacemaker Associated Dysrhythmias
    • Sensor-induced tachycardia
      • Pacemakers designed to respond to physiologic stress by increasing heart rate (i.e. during exercise, hypercapnia, tachypnea)
      • Pacemaker reacts to stimuli not intended to increse heart rate (vibrations, electrocautery, muscle contractions) resulting in inappropriate rate (rate will not exceed pacemaker’s upper rate limit)
      • Treatment: Application of a magnet will terminate
    • Pacemaker mediated tachycardia (PMT) (aka endless loop tachycardia)
      • Re-entry Tachycardia: Ventricular depolarization conducts retrograde into the atria leading atrial lead to detect activity as incoming P wave resulting in ventricular depolarization (viscous cycle develops)
      • Treatment
        • Administer AV blocker (adenosine, beta blocker, calcium channel blocker)
        • Apply magnet over pacemaker

Pacemaker Mediated Tachycardia (learningecg.blogspot.com)

 

Pacemaker Mediated Tachycardia (LITFL)

  • Runaway pacemaker (Ortega 2005)
    • Life-threatening malfunction seen in older-generation pacemakers resulting from low battery voltage
    • Pacemaker delivers runs of pacing spikes in excess of 2000 bp
      • Can provoke ventricular fibrillation
      • Can also cause failure to capture as spikes low amplitude
      • Treatment
        • Apply magnet over pacemaker
        • Replace pacemaker

Runaway Pacemaker (LITFL)

  • Lead displacement dysrhythmia
    • Results from a dislodged pacing wire “floating” around the right ventricle
    • EKG: May demonstrate changing QRS morphology
    • CXR: May aid in diagnosis
  • Twiddler’s Syndrome
    • Dysfunction of the pacemaker resulting from patient manipulation of the pulse generator
    • Repeated manipulation results in the pacemaker rotating on it’s axis and dislodgment of the pacing leads
    • Can cause pacing of the diaphragm or brachial plexus (arm twitching)

Twiddler’s Syndrome (NEJM)

For More on This Topic Checkout:

References:

  1. Riemann JT, Squire B: Implantable Cardiac Devices, in Marx JA, Hockberger RS, Walls RM, et al (eds): Rosen’s Emergency Medicine: Concepts and Clinical Practice, ed 8. St. Louis, Mosby, Inc., 2010, (Ch) 80: p 1064-75.
  2. Kucher N. Deep-vein thrombosis of the upper extremities. NEJM 2011; 364:861. PMID: 21366477
  3. Jacob S et al. Cardiac rhtyhm device identification algorithm using X-Rays: CaRDIA-X. Heart Rhythm 2011; 8(6): 915-22. PMID: 21220049
  4. Ortega DF. Runaway pacemaker: A forgotten phenomenon? Europace 2005; 7(6): 592-7. PMID: 16216762

Post Peer Reviewed By: Salim R. Rezaie (Twitter: @srrezaie)

Share This:

Leave a Comment

Your email address will not be published. Required fields are marked *

emDOCs subscribes to the Free Open Access Meducation (FOAMed) initiative. Our goal is to inform the global EM community with timely and high-yield content about what providers like YOU are seeing and doing daily in your local ED.

WRITE FOR EMDOCS

We are actively recruiting both new topics and authors.
This project is rolling and you can submit an idea or write-up anytime!
Contact us at editors@emdocs.net

news, headlines, newsletter

Join our Newsletter

Keep up to date on all of the latest new articles, studies, and Podcasts.
Overcoming loneliness : strategies for a fulfilling life | myrna young life coach.